DM 1 - DM 1G

Pre-heating furnace for removable prostheses, ideal for medium and large-sized laboratories

Advantages

  • Superior safety, quality and precision with respect to traditional gas furnaces
  • Extremely safe even during the pre-heating phase of alcohol-based coatings
  • The size and structure of the chambers and the heating power allow greater load support
  • 1-year warranty

Characteristics

  • DM 1 (12 cylinders - 9X)
  • DM 1 G (24 cylinders - 9X)
  • Reinforced thermocouple
  • Microswitch detects the opening of the hatch, interrupting the power supply to the resistor
  • Preset extractor command to allow release of fumes
  • Possibility of memorising up to 10 different cycles (programs), each made up of a maximum of 4 positions
  • Temperature control carried out with microprocessor electronics, allows the user to set the following parameters for each program:
    • temperature
    • maintenance temperature
    • Speed at which temperature rises (expressed in °C/min)
    • extractor OFF temperature 
  • Automatic management of the extractor depending on program
  • Delayed start with the possibility of setting the day and time of furnace activity
  • Maximum working temperature 1100°C
  • Supplied with the extractor (2 extractor in the DM 1 G)

 

C305314 DM 1 G    Discontinued product, available while supplies last.
C305313 DM 1        Discontinued product, available while supplies last.

Technical Data

DM 1
Size
L61 cm/ D65 cm/ H82 cm
Internal chamber sizeL30 cm /D23 cm/H18 cm
Cylinder capacity
12 - 9X
Total power
5000 Watt
Voltage (V-Hz)230 V ± 10% / 50-60 Hz
Weight
90 Kg
Max. Heat. Speed from 0 to 900°C60:00 min

DM 1 G
Size L61cm / D82 cm/ H82 cm
Internal chamber sizeL30 cm /D40 cm /H18 cm
Cylinder capacity
24 - 9X
Total power (Watt)7500 Watt
Voltage  (V-Hz)380V ± 10% / 50-60 Hz
Weight 120 Kg
Max. Heat. Speed from 0 to 900°C52:00 min

Packaging

C305313DM 1 pre-heating furnace
C305314DM 1 G pre-heating furnace

Standard components

DM 1 single ectractor
DM 1 G twin extractor

Troubleshooting

ISSUE
CAUSE
REMEDY
After powering the equipment via
the main switch the on-key LED fails to come on.
1. Plug is not connected to the
mains socket.
Power lead is damaged.
2. Fuse blown.
1. Connect the plug to the mains socket
properly. Have the power lead replaced
by an Authorised Service Centre only.
2. Change the fuse, using one of identical
characteristics. Contact an
Authorised
Service Centre
only.
Insufficient aspiration of
combustion fumes and gases.
1. Aspirating fan full of dust
or clogged.
1. Clean the aspiration fan or have it changed.
Contact an
Authorised Service Centre only
Combustion fumes or gas exits
aspiration fan flue or the mobile hood.
1. Obstruction in the oven
aspiration fan flue or mobile
hood is clogged.
1. Clear any obstruction from flue.
The legend “Door” appears on the
display.
1. The door has accidentally
been opened.
1. Re-close the door.
The buzzer sounds and the legend
“tEmP” flashes on the display.
1. Temperature setting exceeded
(see section 7.2 point
B.).
2. Thermocouple broken.
1. Reset the work cycle data.
2. Change the thermocouple.
Contact an
Authorised Service Centre only.
To read all the documents in PDF format, you need to download a PDF viewer program. We suggest Adobe® Acrobat® Reader program, available for free from the following web site: www.adobe.com Note: programs versions are available for all operating systems.

© Zhermack SpA - All rights reserved - Via Bovazecchino, 100 45021 Badia Polesine (RO), Italy
Tel. +39 0425 597611 - Fax. +39 0425 53596 - Email web@zhermack.com
P.IVA 00594630295 - C.F. P.IVA IT00594630295 - R.E.A. 86603 - R.I. RO 00594630295 - M. RO 001381Isc. Reg. A.E.E. IT07111000002 - Cap. Soc. € 1.032.000,00

Credits - Terms and Conditions of Use - Work with us